Designs for an asymmetric gradient set and a compact superconducting magnet for neural magnetic resonance imaging
Data(s) |
01/01/1999
|
---|---|
Resumo |
Imaging of the head and neck is the most commonly performed clinical magnetic resonance imaging (MRI) examination [R. G. Evans and J. R. G. Evans, AJR 157, 603 (1991)]. This is usually undertaken in a generalist MRI instrument containing superconducting magnet system capable of imaging all organs. These generalist instruments are large, typically having a bore of 0.9-1.0 m and a length of 1.7-2.5 m and therefore are expensive to site, somewhat claustrophobic to the patient, and offer little access by attending physicians. In this article, we present the design of a compact, superconducting MRI magnet for head and neck imaging that is less than 0.8 m in length and discuss in detail the design of an asymmetric gradient coil set, tailored to the magnet profile. In particular, the introduction of a radio-frequency FM modulation scheme in concert with a gradient sequence allows the epoch of the linear region of the gradient set to be much closer to the end of the gradient structure than was previously possible. Images from a prototype gradient set demonstrate the effectiveness of the designs. (C) 1999 American Institute of Physics. [S0034-6748(99)04910-2]. |
Identificador | |
Idioma(s) |
eng |
Publicador |
American Institute of Physics |
Palavras-Chave | #Instruments & Instrumentation #Physics, Applied #Coil Design #Optimization #Modulation #Shifts #C1 #671402 Medical instrumentation #291599 Biomedical Engineering not elsewhere classified |
Tipo |
Journal Article |