Resumo

The aim of this work is to propose a biomonitoring method for the simultaneous determination of Cd and Pb in whole blood by simultaneous electrothermal atomic absorption spectrometry for assessment of environmental levels. A volume of 200 mu L of whole blood was diluted in 500 mu L of 0.2% (w v(-1)) Triton(R) X-100 + 2.0% (v v(-1)) HNO3. Trichloroacetic acid was added for protein precipitation and the supernatant analyzed. A mixture of 250 mu g W + 200 mu g Rh as permanent and 2.0% (w v(-1)) NH4H2PO4 as co-injected modifiers were used. Characteristic masses and limits of detections (n = 20, 3s) for Cd and Pb were 1.26 and 33 pg and 0.026 mu g L-1 and 0.65 mu g L-1, respectively. Repeatability ranged from 1.8 to 6.8% for Cd and 1.2 to 1.7% for Pb. The trueness of method was checked by the analysis of three Reference Materials: Lyphocheck(R) Whole Blood Metals Control level 1 and Seronorm(TM) Trace Elements in Whole Blood levels 1 and 2. The found concentrations presented no statistical differences at the 95% confidence level. Blood samples from 40 volunteers without occupational exposure were analyzed and the concentrations ranged from 0.13 to 0.71 mu g L-1 (0.32 +/- 0.19 mu g L-1) for Cd and 9.3 to 56.7 mu g L-1 (25.1 +/- 10.8 mu g L-1) for Pb. (C) 2007 Elsevier B.V. All rights reserved.

Identificador

TALANTA, v.75, n.1, p.246-252, 2008

0039-9140

http://producao.usp.br/handle/BDPI/19894

10.1016/j.talanta.2007.11.003

http://dx.doi.org/10.1016/j.talanta.2007.11.003

Idioma(s)

eng

Publicador

ELSEVIER SCIENCE BV

Relação

Talanta

Direitos

restrictedAccess

Copyright ELSEVIER SCIENCE BV

Palavras-Chave #atomic absorption spectrometry #simultaneous determination #biomonitoring #blood #cadmium #lead #BIOLOGICAL SAMPLES #REFERENCE VALUES #ZEEMAN CORRECTION #TRACE-ELEMENTS #MERCURY #PB #TOXICITY #URINE #CD #MODIFIER #Chemistry, Analytical
Tipo

article

original article

publishedVersion