Mock circulatory system for the evaluation of left ventricular assist devices, endoluminal prostheses, and vascular diseases
Contribuinte(s) |
UNIVERSIDADE DE SÃO PAULO |
---|---|
Data(s) |
18/10/2012
18/10/2012
2008
|
Resumo |
A new digital computer mock circulatory system has been developed in order to replicate the physiologic and pathophysiologic characteristics of the human cardiovascular system. The computer performs the acquisition of pressure, flow, and temperature in an open loop system. A computer program has been developed in Labview programing environment to evaluate all these physical parameters. The acquisition system was composed of pressure, flow, and temperature sensors and also signal conditioning modules. In this study, some results of flow, cardiac frequencies, pressures, and temperature were evaluated according to physiologic ventricular states. The results were compared with literature data. In further works, performance investigations will be conducted on a ventricular assist device and endoprosthesis. Also, this device should allow for evaluation of several kinds of vascular diseases. |
Identificador |
ARTIFICIAL ORGANS, v.32, n.6, p.461-467, 2008 0160-564X http://producao.usp.br/handle/BDPI/18746 10.1111/j.1525-1594.2008.00569.x |
Idioma(s) |
eng |
Publicador |
WILEY-BLACKWELL |
Relação |
Artificial Organs |
Direitos |
restrictedAccess Copyright WILEY-BLACKWELL |
Palavras-Chave | #hemodynamic performance #mock circulatory system #pulsatile blood flow #in vitro experiment #ARTIFICIAL-HEART ATAH #DESIGN #Engineering, Biomedical #Transplantation |
Tipo |
article proceedings paper publishedVersion |