Sequential Injection Analysis (SIA) for Arsenic Speciation by Capillary Electrophoresis Hyphenated to Inductively Coupled Plasma Sector Field Mass Spectrometry (CE-ICP-SFMS)


Autoria(s): SUAREZ, Carlos A.; ARAUJO, Georgia C. L.; GINE, Maria F.; KAKAZU, Mauricio H.; SARKIS, Jorge E. S.
Contribuinte(s)

UNIVERSIDADE DE SÃO PAULO

Data(s)

18/10/2012

18/10/2012

2009

Resumo

Sequential injection analysis (SIA) is proposed for managing microvolumes of sample and arsenic species solutions for speciation analysis by capillary electrophoresis focusing on the reduction of hazardous waste residues. An electronically controlled hydrodynamic injector was projected to introduce microvolumes of solutions prepared by SIA into the CE capillary with precision better than 2%. The determination of arsenite, arsenate, monomethylarsonic acid, dimethylarsinic acid, and arsenobetaine was performed from 50 mu L volumes of lyophilized urine and extract of shrimp with the system hyphenated to inductively coupled plasma mass spectrometry (CE-ICP-SFMS).

CNPq

FINEP

Identificador

SPECTROSCOPY LETTERS, v.42, n.6/Jul, p.376-382, 2009

0038-7010

http://producao.usp.br/handle/BDPI/17180

10.1080/00387010903187286

http://dx.doi.org/10.1080/00387010903187286

Idioma(s)

eng

Publicador

TAYLOR & FRANCIS INC

Relação

Spectroscopy Letters

Direitos

restrictedAccess

Copyright TAYLOR & FRANCIS INC

Palavras-Chave #arsenic species #capillary electrophoresis #hazardous waste #inductively coupled plasma mass spectrometry #sequential injection analysis #AUTOMATED SAMPLE PREPARATION #HYDRIDE GENERATION #WATER SAMPLES #SYSTEM #SEPARATOR #Spectroscopy
Tipo

article

original article