A flow-based analytical procedure for salbutamol determination exploiting chemiluminescence in a liquid-core waveguide
| Contribuinte(s) |
UNIVERSIDADE DE SÃO PAULO |
|---|---|
| Data(s) |
18/10/2012
18/10/2012
2008
|
| Resumo |
Salbutamol is a bronchodilator whose use is restricted due to its anabolic effects. A flow-based procedure for salbutamol determination based on the inhibition of chemiluminescence of the luminol/hypochlorite system was developed. A flow cell constructed with a liquid-core waveguide was employed to constrain the emitted radiation, minimizing losses during transport to detector. Linear response was observed within 2.5 x 10(-6) and 1.0 x 10(-5) mol L-1 with a detection limit estimated as 1 x 10(-7) mol L-1 at the 99.7% confidence level. The coefficient of variation (n = 20), sampling rate, and luminol consumption per determination were estimated as 2.8%, 164 determinations h(-1), and 50 mu g, respectively. Results for pharmaceutical samples were in agreement with those obtained by reference procedures at the 95% confidence level. |
| Identificador |
ANALYTICAL LETTERS, v.41, n.9, p.1579-1591, 2008 0003-2719 http://producao.usp.br/handle/BDPI/17067 10.1080/00032710802122156 |
| Idioma(s) |
eng |
| Publicador |
TAYLOR & FRANCIS INC |
| Relação |
Analytical Letters |
| Direitos |
restrictedAccess Copyright TAYLOR & FRANCIS INC |
| Palavras-Chave | #chemiluminescence #FIA #hypochlorite #liquid-core waveguide #salbutamol #INJECTION-ANALYSIS #CAPILLARY-ELECTROPHORESIS #CHLORINE #REAGENT #Chemistry, Analytical |
| Tipo |
article proceedings paper publishedVersion |