Multifilter rotating shadowband radiometer calibration for spectral aerosol optical depth retrievals over Sao Paulo City, Brazil
Contribuinte(s) |
UNIVERSIDADE DE SÃO PAULO |
---|---|
Data(s) |
18/04/2012
18/04/2012
2008
|
Resumo |
Multifilter rotating shadowband radiometer (MFRSR) calibration values for aerosol optical depth (AOD) retrievals were determined by means of the general method formulated by Forgan [Appl. Opt. 33, 4841 (1994)] at a polluted urban site. The obtained precision is comparable with the classical method, the Langley plot, applied on clean mountaintops distant of pollution sources. The AOD retrieved over Sao Paulo City with both calibration procedures is compared with the Aerosol Robotic Network data. The observed results are similar, and, except for the shortest wavelength (415 nm), the MFRSR`s AOD is systematically overestimated by similar to 0.03. (c) 2008 Optical Society of America. |
Identificador |
APPLIED OPTICS, v.47, n.9, p.1171-1176, 2008 0003-6935 http://producao.usp.br/handle/BDPI/15577 10.1364/AO.47.001171 |
Idioma(s) |
eng |
Publicador |
OPTICAL SOC AMER |
Relação |
Applied Optics |
Direitos |
closedAccess Copyright OPTICAL SOC AMER |
Palavras-Chave | #INSTRUMENT #Optics |
Tipo |
article original article publishedVersion |