Construção de um calorímetro isotérmico diferencial de alta sensibilidade e baixo custo


Autoria(s): Trinca, Rafael Bergamo; Perles, Carlos Eduardo; Volpe, Pedro Luiz Onófrio
Contribuinte(s)

Universidade Estadual de Campinas

Data(s)

01/01/2009

03/12/2015

03/12/2015

Resumo

The high cost of sensitivity commercial calorimeters may represent an obstacle for many calorimetric research groups. This work describes the construction and calibration of a batch differential heat conduction calorimeter with sample cells volumes of about 400 μL. The calorimeter was built using two small high sensibility square Peltier thermoelectric sensors and the total cost was estimated to be about US$ 500. The calorimeter was used to study the excess enthalpy of solution of binary mixtures of liquids, as a function of composition, for the following binary systems of solvents: water + 1,4-dioxane or + dimethylsulfoxide at 298,2 ± 0,5 K.

1651

1654

Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)

Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

Identificador

Química Nova. Sociedade Brasileira de Química, v. 32, n. 6, p. 1651-1654, 2009.

0100-4042

S0100-40422009000600050

10.1590/S0100-40422009000600050

http://dx.doi.org/10.1590/S0100-40422009000600050

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422009000600050

http://www.repositorio.unicamp.br/jspui/handle/REPOSIP/23859

http://repositorio.unicamp.br/jspui/handle/REPOSIP/202478

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Relação

Química Nova

Direitos

aberto

Fonte

SciELO

Palavras-Chave #calorimeter #excess enthalpy of solution #binary mixture
Tipo

Artigo de periódico