Utilização de um monitor de vídeo como fonte de alta tensão para eletroforese capilar


Autoria(s): Silva, José Alberto Fracassi da; Frasatto, Leonardo
Contribuinte(s)

Universidade Estadual de Campinas

Data(s)

01/12/2006

03/12/2015

03/12/2015

Resumo

This article describes the use of a conventional CRT monitor as a high voltage power supply for capillary electrophoresis. With this monitor, a 23-kV high voltage with a ripple of 1.32% was observed. The reproducibility of the applied high voltage was evaluated by measuring the standard deviations of peak area and migration time for five consecutive injections of a test mixture containing potassium, sodium, and lithium cations at 50 mmol L-1. The errors were about 2.5% and 0.6% for peak area and migration time, respectively. The maximum current tested was about 180 mA, which covers most capillary electrophoresis applications. This system has been successfully used for several months, maintaining the desired level of performance.

1377

1380

Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)

Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

Identificador

Química Nova. Sociedade Brasileira de Química, v. 29, n. 6, p. 1377-1380, 2006.

0100-4042

S0100-40422006000600038

10.1590/S0100-40422006000600038

http://dx.doi.org/10.1590/S0100-40422006000600038

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422006000600038

http://www.repositorio.unicamp.br/jspui/handle/REPOSIP/23733

http://repositorio.unicamp.br/jspui/handle/REPOSIP/202426

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Relação

Química Nova

Direitos

aberto

Fonte

SciELO

Palavras-Chave #capillary electrophoresis #high voltage power supply #analytical instrumentation
Tipo

Artigo de periódico