Utilização de um monitor de vídeo como fonte de alta tensão para eletroforese capilar
Contribuinte(s) |
Universidade Estadual de Campinas |
---|---|
Data(s) |
01/12/2006
03/12/2015
03/12/2015
|
Resumo |
This article describes the use of a conventional CRT monitor as a high voltage power supply for capillary electrophoresis. With this monitor, a 23-kV high voltage with a ripple of 1.32% was observed. The reproducibility of the applied high voltage was evaluated by measuring the standard deviations of peak area and migration time for five consecutive injections of a test mixture containing potassium, sodium, and lithium cations at 50 mmol L-1. The errors were about 2.5% and 0.6% for peak area and migration time, respectively. The maximum current tested was about 180 mA, which covers most capillary electrophoresis applications. This system has been successfully used for several months, maintaining the desired level of performance. 1377 1380 Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP) Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq) |
Identificador |
Química Nova. Sociedade Brasileira de Química, v. 29, n. 6, p. 1377-1380, 2006. 0100-4042 S0100-40422006000600038 10.1590/S0100-40422006000600038 http://dx.doi.org/10.1590/S0100-40422006000600038 http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422006000600038 http://www.repositorio.unicamp.br/jspui/handle/REPOSIP/23733 |
Idioma(s) |
pt |
Publicador |
Sociedade Brasileira de Química |
Relação |
Química Nova |
Direitos |
aberto |
Fonte |
SciELO |
Palavras-Chave | #capillary electrophoresis #high voltage power supply #analytical instrumentation |
Tipo |
Artigo de periódico |