Metodologia analítica para determinação de folatos e ácido fólico em alimentos
Contribuinte(s) |
Universidade Estadual de Campinas |
---|---|
Data(s) |
01/10/2006
03/12/2015
03/12/2015
|
Resumo |
The aim of this work was to develop and to validate a methodology using HPLC for the simultaneous determination of folates and folic acid in foods. The limits of detection and the recovery rates for the vitamins in the certified reference materials were respectively 5 pg/mL and 94-108% for 5-MTHF, 7 pg/mL and 97-102% for THF, 30 pg/mL and 97.9-104% for 5-FTHF, 30 pg/mL and 95-107 for 10-FFA, 5 ng/mL and 97-102% for FA and 5 ng/mL and 98-103% for 10-MFA. Repeatability showed a coefficient of variation below 3.9% for all the vitamins. The proposed methodology was shown to be efficient when applied to different certified reference materials, namely pig's liver (BCR487), powdered milk (BCR421) and a vegetable mixture (BCR485). 972 976 |
Identificador |
Química Nova. Sociedade Brasileira de Química, v. 29, n. 5, p. 972-976, 2006. 0100-4042 S0100-40422006000500016 10.1590/S0100-40422006000500016 http://dx.doi.org/10.1590/S0100-40422006000500016 http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422006000500016 http://www.repositorio.unicamp.br/jspui/handle/REPOSIP/23724 |
Idioma(s) |
pt |
Publicador |
Sociedade Brasileira de Química |
Relação |
Química Nova |
Direitos |
aberto |
Fonte |
SciELO |
Palavras-Chave | #HPLC #methodology #folates #food |
Tipo |
Artigo de periódico |