On the assessment of time-shift variations from backscattered ultrasound for large temperature changes in biological phantoms
Contribuinte(s) |
Garreton, L. G. |
---|---|
Data(s) |
13/04/2015
13/04/2015
01/01/2010
|
Resumo |
This work reports the assessment of time-shifts (TS) from backscattered ultrasound (BSU) signals when large temperature variations (up to 15 degrees C) were induced in a gel-based phantom. The results showed that during cooling temperature is linear with TS at a rate of approximately 74 ns/degrees C. However during a complete heating/cooling cycle, the relation is highly non-linear. This can be explained by the fact that during cooling the temperature distribution is more uniform. Another problem to report is that TS is very sensitive to external movements. |
Identificador |
1875-3892 AUT: MRU00118; |
Idioma(s) |
eng |
Publicador |
Elsevier |
Relação |
Physics Procedia |
Palavras-Chave | #Non-invasive temperature estimation #Temperature models #Time-shift estimation #Ultrasound |
Tipo |
conferenceObject |
Direitos |
restrictedAccess |