The use of Log-file analysis within VMAT audits


Autoria(s): McGarry, Conor K; Agnew, Christina E; Hussein, Mohammad; Tsang, Yatman; Hounsell, Alan R; Clark, Catharine H
Data(s)

2016

Resumo

<p>OBJECTIVE: This work investigates the delivery accuracy of different Varian linear accelerator models using log-file derived MLC RMS values.</p><p>METHODS: Seven centres independently created a plan on the same virtual phantom using their own planning system and the log files were analysed following delivery of the plan in each centre to assess MLC positioning accuracy. A single standard plan was also delivered by seven centres to remove variations in complexity and the log files were analysed for Varian TrueBeams and Clinacs (2300IX or 2100CD models).</p><p>RESULTS: Varian TrueBeam accelerators had better MLC positioning accuracy (<1.0mm) than the 2300IX (<2.5mm) following delivery of the plans created by each centre and also the standard plan. In one case log files provided evidence that reduced delivery accuracy was not associated with the linear accelerator model but was due to planning issues.</p><p>CONCLUSIONS: Log files are useful in identifying differences between linear accelerator models, and isolate errors during end-to-end testing in VMAT audits. Log file analysis can rapidly eliminate the machine delivery from the process and divert attention with confidence to other aspects. Advances in Knowledge: Log file evaluation was shown to be an effective method to rapidly verify satisfactory treatment delivery when a dosimetric evaluation fails during end-to-end dosimetry audits. MLC RMS values for Varian TrueBeams were shown to be much smaller than Varian Clinacs for VMAT deliveries.</p>

Identificador

http://pure.qub.ac.uk/portal/en/publications/the-use-of-logfile-analysis-within-vmat-audits(7b56c48d-6056-41db-9169-03736da04400).html

http://dx.doi.org/10.1259/bjr.20150489

Idioma(s)

eng

Direitos

info:eu-repo/semantics/closedAccess

Fonte

McGarry , C K , Agnew , C E , Hussein , M , Tsang , Y , Hounsell , A R & Clark , C H 2016 , ' The use of Log-file analysis within VMAT audits ' British Journal of Radiology , vol 89 , no. 1062 , 20150489 . DOI: 10.1259/bjr.20150489

Tipo

article