Application of a cw quantum cascade laser CO2 analyser to catalytic oxidation reaction monitoring


Autoria(s): Kasyutich, V. L.; Poulidi, D.; Jalil, M.; Metcalfe, I. S.; Martin, P. A.
Data(s)

01/02/2013

Resumo

<p>Catalytic oxidation reaction monitoring has been performed for the first time with a trace gas carbon dioxide analyser based on a continuous wave (cw), thermoelectrically cooled (TEC), distributed feedback (DFB) quantum cascade laser (QCL) operating at around 2307 cm<sup>-1</sup>. The reaction kinetics for carbon monoxide oxidation over a platinum catalyst supported on yttria-stabilised zirconia were followed by the QCL CO<sub>2</sub> analyser and showed that it is a powerful new tool for measuring low reaction rates associated with low surface area model catalysts operating at atmospheric pressures. A detection limit was determined of 40 ppb (1 standard deviation) for a 0.1 s average and a residual absorption standard deviation of 1.9×10<sup>-4</sup>. © 2012 Springer-Verlag.</p>

Identificador

http://pure.qub.ac.uk/portal/en/publications/application-of-a-cw-quantum-cascade-laser-co2-analyser-to-catalytic-oxidation-reaction-monitoring(8155a942-0842-4997-b9fa-f99df78c73ad).html

http://dx.doi.org/10.1007/s00340-012-5154-y

http://www.scopus.com/inward/record.url?scp=84883488580&partnerID=8YFLogxK

Idioma(s)

eng

Direitos

info:eu-repo/semantics/restrictedAccess

Fonte

Kasyutich , V L , Poulidi , D , Jalil , M , Metcalfe , I S & Martin , P A 2013 , ' Application of a cw quantum cascade laser CO2 analyser to catalytic oxidation reaction monitoring ' Applied Physics B - Lasers and Optics , vol 110 , no. 2 , pp. 263-269 . DOI: 10.1007/s00340-012-5154-y

Palavras-Chave #/dk/atira/pure/subjectarea/asjc/3100 #Physics and Astronomy(all) #/dk/atira/pure/subjectarea/asjc/3100/3101 #Physics and Astronomy (miscellaneous)
Tipo

article