In vitro fluid dynamics of the Ahmed glaucoma valve modified with expanded polytetrafluoroethylene.


Autoria(s): DeCroos, FC; Kondo, Y; Mordes, D; Lee, MR; Ahmad, S; Asrani, S; Allingham, RR; Olbrich, KC; Klitzman, B
Data(s)

01/02/2011

Formato

112 - 117

Identificador

http://www.ncbi.nlm.nih.gov/pubmed/21281065

Curr Eye Res, 2011, 36 (2), pp. 112 - 117

http://hdl.handle.net/10161/10348

1460-2202

Relação

Curr Eye Res

10.3109/02713683.2010.512115

Palavras-Chave #Gelatin #Glaucoma Drainage Implants #Hydrodynamics #In Vitro Techniques #Polytetrafluoroethylene #Pressure #Sodium Chloride #Transducers, Pressure
Tipo

Journal Article

Cobertura

England

Resumo

PURPOSE: Long-term intraocular pressure reduction by glaucoma drainage devices (GDDs) is often limited by the fibrotic capsule that forms around them. Prior work demonstrates that modifying a GDD with a porous membrane promotes a vascularized and more permeable capsule. This work examines the in vitro fluid dynamics of the Ahmed valve after enclosing the outflow tract with a porous membrane of expanded polytetrafluoroethylene (ePTFE). MATERIALS AND METHODS: The control and modified Ahmed implants (termed porous retrofitted implant with modified enclosure or PRIME-Ahmed) were submerged in saline and gelatin and perfused in a system that monitored flow (Q) and pressure (P). Flow rates of 1-50 μl/min were applied and steady state pressure recorded. Resistance was calculated by dividing pressure by flow. RESULTS: Modifying the Ahmed valve implant outflow with expanded ePTFE increased pressure and resistance. Pressure at a flow of 2 μl/min was increased in the PRIME-Ahmed (11.6 ± 1.5 mm Hg) relative to the control implant (6.5 ± 1.2 mm Hg). Resistance at a flow of 2 μl/min was increased in the PRIME-Ahmed (5.8 ± 0.8 mm Hg/μl/min) when compared to the control implant (3.2 ± 0.6 mm Hg/μl/min). CONCLUSIONS: Modifying the outflow tract of the Ahmed valve with a porous membrane adds resistance that decreases with increasing flow. The Ahmed valve implant behaves as a variable resistor. It is partially open at low pressures and provides reduced resistance at physiologic flow rates.

Idioma(s)

ENG