Capillary electrochromatography with monolithic poly(styrene-co-divinylbenzene-co-methacrylic acid) as the stationary phase


Autoria(s): Xiong, BH; Zhang, LH; Zhang, YK; Zou, HF; Wang, JD
Data(s)

2000

Resumo

A new kind of monolithic capillary electrochromatography column with poly(styrene-co-divinylbenzene-co-methacrylic acid) as the stationary phase has been developed. The stationary phase was found to be porous by scanning electron microscopy and the composition of the continuous bed was proved by IR spectroscopy to be the ternary polymer of styrene, divinylbenzene, and methacrylic acid. The effects of operating parameters, such as voltage, electrolyte, and organic modifier concentration in the mobile phase on electroosmotic flow were studied systematically, The retention mechanism of neutral solutes on such a column proved to be similar to that of reversed-phase high performance liquid chromatography. In addition, fast analyses of phenols, chlorobenzenes, anilines, isomeric compounds of phenylenediamine and alkylbenzenes within 4.5 min were achieved.

Identificador

http://159.226.238.44/handle/321008/91455

http://www.irgrid.ac.cn/handle/1471x/183739

Idioma(s)

英语

Fonte

熊博晖;张丽华;张玉奎; 邹汉法;王俊德.Capillary Electrochromatography with Monolithic Poly(styrene-co-divinylbenzene-co-methacrylic acid) as the Stationary Phase,HRC-Journal of High Resolution Chromatography,2000,1(23):67-72

Palavras-Chave #capillary electrochromatography #monolithic column #electroosmotic flow #retention mechanism #aromatic compound
Tipo

期刊论文