AMPEROMETRIC DETECTION AT A CARBON-FIBER ARRAY RING GLASSY-CARBON DISK ELECTRODE IN A WALL-JET CELL
Data(s) |
1991
|
---|---|
Resumo |
A wall-jet cell incorporating a carbon fibre array ring/glassy-carbon disk electrode has been constructed, and characterized by the cyclic voltammetry and flow-injection techniques. The ring (composed of several microdisks) and glassy-carbon disk electrode, can be used separately for different purposes, e.g., detection in solution without a supporting electrolyte, collection/shielding detection with dual-electrode and voltammetric/amperometric detection with series dual-electrode. The electrode shows better collection and shielding effects than usual ring-disk electrode in quiescent solution and the series dual-electrode in a thin-layer flow-through cell. The detection limit at the ring electrode is comparable with that at a conventional-size electrode, and has been used in the mobile phase without a supporting electrolyte, proving to be a promising detector for normal-phase liquid chromatography. |
Identificador | |
Idioma(s) |
英语 |
Fonte |
JI HM;WANG E.AMPEROMETRIC DETECTION AT A CARBON-FIBER ARRAY RING GLASSY-CARBON DISK ELECTRODE IN A WALL-JET CELL,TALANTA,1991,38(1):73-80 |
Palavras-Chave | #PERFORMANCE LIQUID-CHROMATOGRAPHY #COLUMN ELECTROCHEMICAL DETECTOR #CHEMICALLY MODIFIED ELECTRODE #VOLTAMMETRIC DETECTOR #HUMAN-SERUM #MICROELECTRODES #SYSTEMS #SENSORS #CHANNEL #SINGLE |
Tipo |
期刊论文 |