AMPEROMETRIC DETECTION OF AMINO-ACIDS IN A FLOW-INJECTION SYSTEM WITH A NICKEL(II)-MODIFIED ELECTRODE WITH AN EASTMAN-AQ POLYMER FILM


Autoria(s): LIU AH; WANG EK
Data(s)

1993

Resumo

Amperometic flow measurements were made at +0.55 V (vs. Ag/AgCl) in 0.1 mol l-1 KOH electrolyte with an Ni(II) chemically modified electrode (CME) with an Eastman-AQ polymer film. The use and characteristics of a Ni(II)-containing crystalline and polymer-modified electrode obtained by a double coating step as a detector for amino acids in a flow-injection system using reversed-phase liquid chromatography are described. The detection of these analytes is based on the higher oxidation state of nickel (NiOOH) controlled by the applied potential. The electroanalytical parameters and the detection current for a series of amines and amino acids were investigated. The use of such a CME in the flow-injection technique was found to be suitable in a solution at low pH. The linear range for glycine is 5 X 10(-6)-0.1 mol 1-1 with a detection limit of 1.0 X 10(-6) mol l-1. A 1 X 10(-4) mol 1-1 mixture of serine and tyrosine was also detected after separation on an Nucleosil C18 column.

Identificador

http://ir.ciac.jl.cn/handle/322003/36065

http://www.irgrid.ac.cn/handle/1471x/161555

Idioma(s)

英语

Fonte

LIU AH;WANG EK.AMPEROMETRIC DETECTION OF AMINO-ACIDS IN A FLOW-INJECTION SYSTEM WITH A NICKEL(II)-MODIFIED ELECTRODE WITH AN EASTMAN-AQ POLYMER FILM,ANALYTICA CHIMICA ACTA,1993,280(2):223-229

Palavras-Chave #PERFORMANCE LIQUID-CHROMATOGRAPHY #CHEMICALLY MODIFIED ELECTRODE #NICKEL-OXIDE ELECTRODE #ELECTROCHEMICAL DETECTION #PEPTIDES #ALDITOLS #SUGARS
Tipo

期刊论文