Detection of menadione sodium bisulfite (vitamin K-3) by reversed-phase high performance liquid chromatography with series dual-electrode amperometric detector


Autoria(s): Liu ZM; Li T; Li J; Wang EK
Data(s)

1997

Resumo

A reversed-phase high-performance liquid chromatography with series dual glassy carbon electrodes for the amperometric detection of water-soluble menadione is described. The complex post-column derivatization reaction and the high background currents were avoided. The menadione sodium bisulfite was reduced at -0.3 V vs. SCE at the upstream (generator) electrode and oxidized at +0.2V vs. SCE at the downstream (collector) electrode. The mobile phase was 0.2moll(-1) HAc-NaAc aqueous buffer (pH 5.50) and 40% (v/v) methanol. The linear response was in the range of 35 ng to 15 mu g, with a detection Limit of 15 ng (S/N=3). The correlation coefficient was 0.9997 (n=6). The electrochemical detection with series dual electrodes has a higher selectivity for menadione (vitamin K-3) compound than with UV detection.

Identificador

http://ir.ciac.jl.cn/handle/322003/24419

http://www.irgrid.ac.cn/handle/1471x/156937

Idioma(s)

英语

Fonte

Liu ZM;Li T;Li J;Wang EK.Detection of menadione sodium bisulfite (vitamin K-3) by reversed-phase high performance liquid chromatography with series dual-electrode amperometric detector,ANALYTICA CHIMICA ACTA,1997,338(1-2 ):57-62

Palavras-Chave #ANIMAL FEED #FLUORIMETRIC DETERMINATION #COLUMN
Tipo

期刊论文