Determination of hydroxylamine by capillary electrophoresis electrochemical detection with a palladium-particle modified carbon fiber microdisk array electrode


Autoria(s): You TY; Wu MJ; Wang EK
Data(s)

1997

Resumo

A highly dispersed ultramicro palladium-particle modified carbon fiber microdisk array electrode (Pd-CFE) was employed for capillary electrophoresis-electrochemical (CEEC) detection of hydroxylamine (HA). The Pd particles obtained were in the nanometer scale, had a high electrocatalytic activity towards HA and exhibited good reproducibility and stability. A linear relationship between the current and the analyte concentration was found between 5 x 10(-6) and 1 x 10(-3) mol/l of HA with a correlation coefficient of 0.9992. The detection limit was 5 x 10(-8) mol/l. The applicability of the method for the determination of HA in river water and waste water was investigated.

Identificador

http://ir.ciac.jl.cn/handle/322003/24343

http://www.irgrid.ac.cn/handle/1471x/156899

Idioma(s)

英语

Fonte

You TY;Wu MJ;Wang EK.Determination of hydroxylamine by capillary electrophoresis electrochemical detection with a palladium-particle modified carbon fiber microdisk array electrode,ANALYTICAL LETTERS,1997,30(5):1025-1036

Palavras-Chave #AMPEROMETRIC DETECTION #MICROELECTRODE
Tipo

期刊论文