A hydrogen peroxide biosensor combined HRP doped polypyrrole with ferrocene modified sol-gel derived composite carbon electrode


Autoria(s): Tian FM; Xu B; Zhu GY
Data(s)

2001

Resumo

A novel amperometric biosensor for the detection of hydrogen peroxide is described. The biosensor was constructed by electrodepositing HRP/PPy membrane on the surface of ferrocenecarboxylic acid mediated sol-gel derived composite carbon electrode. The biosensor gives response to hydrogen peroxide in a few seconds with detection limit of 5x10(-7) mol (.) L-1 (based on signal : noise=3). Linear range is up to 0.2 mmol (.) L-1.

Identificador

http://ir.ciac.jl.cn/handle/322003/20687

http://www.irgrid.ac.cn/handle/1471x/155078

Idioma(s)

英语

Fonte

Tian FM;Xu B;Zhu GY.A hydrogen peroxide biosensor combined HRP doped polypyrrole with ferrocene modified sol-gel derived composite carbon electrode,CHINESE CHEMICAL LETTERS,2001,12(7):625-628

Palavras-Chave #GLUCOSE-OXIDASE ELECTRODE #AMPEROMETRIC DETECTION #CERAMIC-CARBON
Tipo

期刊论文