Real-time immunoassay of antibody activity in serum by surface plasmon resonance biosensor


Autoria(s): Pei RJ; Cui XQ; Yang XR; Wang EK
Data(s)

2000

Resumo

A surface plasmon resonance biosensor has been used to determine antibody activity in serum. As a model system, the interaction of mouse IgG and sheep anti-mouse IgG polyclonal antibody was investigated in real time. The factors, including pH value, ionic strength, protein concentration, influencing electrostatic adsorption of mouse IgG protein onto carboxylated dextran-coated sensor chip surface, were studied. The procedures of mouse IgG protein immobilization and immune reaction were monitored in real time. The regeneration effect using the different elution reagents was also investigated. The same mouse IgG immobilized surface can be used for 100 cycles of binding and elution with only 0.38% loss per regeneration in reactivity. The results show that the surface plasmon resonance biosensor is a rapid, simple, sensitive, accurate and reliable detection technique for real-time immunoassay of antibody activity. The assay allows antibodies to be detected and studied in their native form without any purification. (C) 2000 Elsevier Science B.V. All rights reserved.

Identificador

http://202.98.16.49/handle/322003/19449

http://www.irgrid.ac.cn/handle/1471x/154459

Idioma(s)

英语

Fonte

Pei RJ;Cui XQ;Yang XR;Wang EK.Real-time immunoassay of antibody activity in serum by surface plasmon resonance biosensor,TALANTA,2000,53(3):481-488

Palavras-Chave #BIOSPECIFIC INTERACTION ANALYSIS #COMPLEX #IMMOBILIZATION #AFFINITY #PROTEINS #SENSORS #BINDING #LIGANDS #FILMS #HIV-1
Tipo

期刊论文