Determination of diphenhydramine by capillary electrophoresis with tris(2,2 '-bipyridyl)ruthenium(II) electrochemiluminescence detection
Data(s) |
2003
|
---|---|
Resumo |
Tris(2,2'-bipyridyl)ruthenium(II) electrochemiluminescence detection in a capillary electrophoresis separation system was used for the determination of diphenhydramine. In this study, platinum disk electrode (300 mum in diameter) was used as a working electrode and the influence of applied potential and buffer conditions were investigated. Under optimal conditions: 1.2 V applied potential, pH 8.50, 15 kV separation voltage and 10 mmol l(-1) running buffer, the calibration curve of diphenhydramine was linear over the range of 4 x 10(-8) to 1 x 10(-5) Mol l(-1). This technique gave satisfactory precision, and relative standard deviations of migration times and chemiluminescence peak intensities were less than 1 and 6%, respectively. The technique was applied to animal studies for determination of diphenhydramine extracted from rabbit plasma and urine samples, and the extraction efficiency were between 92 and 98.5%. |
Identificador | |
Idioma(s) |
英语 |
Fonte |
Liu JF;Cao WD;Yang XR;Wang EK.Determination of diphenhydramine by capillary electrophoresis with tris(2,2 '-bipyridyl)ruthenium(II) electrochemiluminescence detection,TALANTA,2003,59(3):453-459 |
Palavras-Chave | #RU(BPY)(3)(3+)-BASED CHEMILUMINESCENCE DETECTION #ELECTROGENERATED CHEMILUMINESCENCE #ELECTROKINETIC CHROMATOGRAPHY #LIQUID-CHROMATOGRAPHY #GAS-CHROMATOGRAPHY #FLOW-INJECTION #AMINO-ACIDS #N-OXIDE #HYDROCHLORIDE #LUMINESCENCE |
Tipo |
期刊论文 |