Determination of benzhexol hydrochloride by capillary zone electrophoresis with an end-column electrochemiluminescence detection


Autoria(s): Yan JL; Liu JF; Cao WD; Sun XH; Yang XR; Wang EK
Data(s)

2004

Resumo

A capillary zone electrophoresis with end-column electrochemiluminescence (ECL) detector was described for the determination of benzhexol hydrochloride. The detection was based on the tris(2,2'-bypyridine)ruthenium(II) [Ru(bpy)(3)(2+)] ECL reaction with the analyte. Electrophoresis was performed using a 25 mum i.d. uncoated capillary. 10 mM sodium phosphate buffer (pH=8.0) was used as the running buffer. The solution in the detection cell was 80 mM sodium phosphate (pH=8.0) and 5 mM)21 Ru(bpy)(3)(2+). A linear calibration curve of three-orders of magnitude was obtained (with a correlation coefficient of > 0.999) from 1.0X10(-8) to 1.0X10(-5) M and the limit of detection was 6.7 X 10(-9) M (S/N= 3). This just provides an easy and sensitive method to determine the active ingredient in pharmaceutical formulations.

Identificador

http://ir.ciac.jl.cn/handle/322003/14839

http://www.irgrid.ac.cn/handle/1471x/150588

Idioma(s)

英语

Fonte

Yan JL;Liu JF;Cao WD;Sun XH;Yang XR;Wang EK.Determination of benzhexol hydrochloride by capillary zone electrophoresis with an end-column electrochemiluminescence detection,MICROCHEMICAL JOURNAL,2004,76(1-2 ):11-16

Palavras-Chave #RU(BPY)(3)(3+)-BASED CHEMILUMINESCENCE DETECTION #IONIZATION MASS-SPECTROMETRY #ELECTROGENERATED CHEMILUMINESCENCE #GAS-CHROMATOGRAPHY #LIQUID-CHROMATOGRAPHY #HUMAN SERUM #TRIHEXYPHENIDYL #DRUGS #TRIS(2 #2'-BIPYRIDYL)RUTHENIUM(III) #PROCYCLIDINE
Tipo

期刊论文