Capillary electrophoresis coupled with electrochemiluminescence detection using porous etched joint


Autoria(s): Yin XB; Qiu HB; Sun XH; Yan JL; Liu JF; Wang EK
Data(s)

2004

Resumo

A new setup to couple capillary electrophoresis (CE) with electrochemiluminescence (ECL) detection is described in which the electrical connection of CE is achieved through a porous section at a distance of 7 mm from the CE capillary outlet. Because the porous capillary wall allowed the CE current to pass through and there was no electric field gradient beyond that section, the influence of CE high-voltage field on the ECL procedure was eliminated. The porous section formed by etching the capillary with hydrofluoric acid after only one side of the circumference of 2-3 mm of polyimide coating of the capillary was removed, while keeping the polyimide coating on the other part to protect the capillary from HF etching makes the capillary joint much more robust since only a part of the circumference of it is etched. A standard three-electrode configuration was used in experiments with Pt wire as a counter electrode, Ag/AgCl as a reference electrode, and a 300-mum diameter Pt disk as a working electrode. Compared with CE-ECL conventional decoupler designs, the present setup with a porous joint has no added dead volume created.

Identificador

http://ir.ciac.jl.cn/handle/322003/14795

http://www.irgrid.ac.cn/handle/1471x/150545

Idioma(s)

英语

Fonte

Yin XB;Qiu HB;Sun XH;Yan JL;Liu JF;Wang EK.Capillary electrophoresis coupled with electrochemiluminescence detection using porous etched joint,ANALYTICAL CHEMISTRY,2004,76(13 ):3846-3850

Palavras-Chave #RU(BPY)(3)(3+)-BASED CHEMILUMINESCENCE DETECTION #COLUMN ELECTROCHEMICAL DETECTION #ZONE-ELECTROPHORESIS #ELECTROGENERATED CHEMILUMINESCENCE #MASS-SPECTROMETRY #INTERFACE #ELECTRODE #PEPTIDES #SENSOR
Tipo

期刊论文