New uniform algorithm to predict reversed phase retention values under different gradient conditions


Autoria(s): Wang, LX; Zhang, LF; Xiao, HB; Liang, XM
Data(s)

10/01/2003

Resumo

A new numerical emulation algorithm was established to calculate retention parameters in RP-HPLC with several retention times under different linear or nonlinear binary gradient elution conditions and further predict the retention time under any other binary gradient conditions. A program was written according to this algorithm and nine solutes were used to test the program. The prediction results were excellent. The maximum relative error of predicted retention time was less than 0.45%. (C) 2002 Elsevier Science B.V. All rights reserved.

Identificador

http://159.226.238.44/handle/321008/82681

http://www.irgrid.ac.cn/handle/1471x/138240

Idioma(s)

英语

Fonte

王龙星; 张乐沣; 肖红斌; 梁鑫淼.New uniform algorithm to predict reversed phase retention values under different gradient conditions,Analytica Chimica Acta,2003,476():123-129

Palavras-Chave #reversed phase HPLC #retention parameter #algorithm #gradient elution
Tipo

期刊论文