Study of an electroosmotic pump for liquid delivery and its application in capillary column liquid chromatography


Autoria(s): Chen, LX; Ma, JP; Guan, YF
Data(s)

05/03/2004

Resumo

A packed-bed electroosmotic pump (EOP) was constructed and evaluated. The EOP consisted of three capillary columns packed in parallel, a gas-releasing device, Pt electrodes and a high-voltage power supply. The EOP could generate output pressure above 5.0 MPa and constant flow rate in the range of nl/min to a few mul/min for pure water, pure methanol, 2 mM potassium dihydrogenphosphate buffer, the buffer-methanol mixture and the pure water-methanol mixture at applied potentials less than 20 W The composition of solvent before/after pumping was quantitatively determined by using a gas chromatograph equipped with both flame ionization detector and thermal conductivity detector. It was found that there were no apparent changes in composition and relative concentrations after pumping process for a methanol-ethanol-acetonitrile mixture and a methanol-water mixture. Theoretical aspect of the EOP was discussed in detail. An capillary HPLC system consisting of the EOP, an injection valve, a 15 cm x 320 mum i.d., 5 mum Spherigel C(18) stainless steel analytical column, and an on-column UV detector was connected to evaluate the performance of the EOP. A comparative study was also carried out with a mechanical capillary HPLC pump on the same system. The results demonstrated that the reproducibility of flow rate and the pulsation-free flow property of the EOP are superior to that of mechanical pump in capillary HPLC application. (C) 2004 Elsevier B.V. All rights reserved.

Identificador

http://159.226.238.44/handle/321008/81699

http://www.irgrid.ac.cn/handle/1471x/136776

Idioma(s)

英语

Fonte

陈令新;马继平;关亚风.Study of an electroosmotic pump for liquid delivery and its application in capillary column liquid chromatography,Journal of Chromatography A,2004,1028():219-226

Palavras-Chave #electroosmotic pump #pumps #instrumentation
Tipo

期刊论文