Physical property measurement of therapeutic carbon ion beam in the shallow-seated tumor therapy terminal at HIRFL


Autoria(s): Dai, ZY (Dai Zhong-Ying); Li, Q (Li Qiang); Yan, Z (Yan Zheng); Jin, XD (Jin Xiao-Dong); Hu, ZG (Hu Zheng-Guo); Xiao, GQ (Xiao Guo-Qing)
Data(s)

3896

Resumo

For the first time the physical properties of therapeutic carbon-ion beam supplied by, the shallow-seated tumor therapy terminal at the Heavy Ion Research Facility in Lanzhou (HIRFL) are measured. For a 80.55MeV/u C-12 ion beam delivered to the therapy terminal, the homogeneity of irradiation fields is 73.48%, when the beam intensity varied in the range of 0.001-0.1nA (i.e. 1 X 10(6) - 1 X 10(8) particles per second). The stability of the beam intensity within a few minutes is estimated to be 80.87%. The depth-dose distribution of the beam at the isocenter of the therapy facility is measured, and the position of the high-dose Bragg peak is found to be located at the water-equivalent depth of 13.866mm. Based on the relationship between beam energy and Bragg peak position, the corresponding beam energy at the isocenter of the therapy terminal is evaluated to be 71.71MeV/u for the original 80.55MeV/u C-12 ion beam, which consisted basically with calculation. The readout of the previously-used air-free ionization chamber regarding absorbed dose is calibrated as well in this experiment. The results indicate that the performance of the therapy facility should be optimized further to meet the requirements of clinical trial.

Identificador

http://ir.impcas.ac.cn/handle/113462/6149

http://www.irgrid.ac.cn/handle/1471x/132760

Idioma(s)

中文

Fonte

Dai, ZY (Dai Zhong-Ying); Li, Q (Li Qiang); Yan, Z (Yan Zheng); Jin, XD (Jin Xiao-Dong); Hu, ZG (Hu Zheng-Guo); Xiao, GQ (Xiao Guo-Qing).Physical property measurement of therapeutic carbon ion beam in the shallow-seated tumor therapy terminal at HIRFL, HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION ,38961,30(9):920-924

Palavras-Chave #heavy ion cancer therapy #homogeneity of irradiation field #stability of beam intensity #depth-dose distribution #Bragg peak
Tipo

期刊论文