Analysis of the practical force accuracy of electromagnet-based nanoindenters


Autoria(s): Huan Y(郇勇); Liu DX(刘东旭); Yang R(杨荣); Zhang TH(张泰华)
Data(s)

2010

Resumo

Two important factors that influence the force accuracy of the electromagnet-based nano-indenters but have not yet attracted much attention are analyzed, and a more reasonable way to estimate the force accuracy is presented in this paper. MTS Nano Indenter (R), with the characteristics of a coil suspended in a uniform magnetic field by two sets of springs acting as an actuator and force measuring unit, is used as an example. One of the two factors is the uniformity of the magnetic field. The other is the stiffness of the supporting spring. Consequently, the practical force accuracy varies considerably from test to test because it firmly depends on the working position of the coil and the displacement stroke. A reasonable estimated accuracy value is of the order of 10 degrees mu N for typical indentation tests with a 10(2) nm indentation depth or a 10 degrees mN test force. (C) 2010 Elsevier Ltd. All rights reserved.

National Natural Science Foundation of China [10602058, 10872200]

Identificador

http://dspace.imech.ac.cn/handle/311007/43467

http://www.irgrid.ac.cn/handle/1471x/124441

Idioma(s)

英语

Fonte

MEASUREMENT.2010,43(9):1090-1093

Palavras-Chave #Nanoindenter #Force #Accuracy #Measurement #Nano-Indentation #Indenter
Tipo

期刊论文