Phase uncertainty in calibrating microwave test fixtures


Autoria(s): Zhu NH
Data(s)

1999

Resumo

The problem of phase uncertainty arising in calibration of the test fixtures is investigated in this paper, It is shown that the problem exists no matter what kinds of calibration standards are used. It is also found that there is no need to determine the individual S-parameters of the test fixtures. In order to eliminate the problem of phase uncertainty, three different precise (known) reflection standards or one known reflection standard plus one known transmission standard should be used to calibrate symmetrical test fixtures. For the asymmetrical cases, three known standards, including at least one transmission standard, should be used. The thru-open-match (TOM) and thru-short-match (TSM) techniques are the simplest methods, and they have no bandwidth limitation. When the standards are imprecise (unknown), it is recommended to use any suitable technique, such as the thru-reflect-line, line-reflect-line, thru-short-delay, thru-open-delay,line-reflect-match, line-reflect-reflect-match, or multiline methods, to accurately determine the values of the required calibration terms and, in addition, to use the TOM or TSM method with the same imprecise standards to resolve the phase uncertainty.

Identificador

http://ir.semi.ac.cn/handle/172111/12792

http://www.irgrid.ac.cn/handle/1471x/65366

Idioma(s)

英语

Fonte

Zhu NH .Phase uncertainty in calibrating microwave test fixtures ,IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES,1999,47(10):1917-1922

Palavras-Chave #光电子学 #calibration #microwave network analyzer #scattering-parameter #measurement #test fixtures #PARAMETERS #NETWORK-ANALYZER CALIBRATION
Tipo

期刊论文