Design, simulation and testing of large area silicon drift detectors and detector array for X-ray spectroscopy
Data(s) |
2000
|
---|---|
Resumo |
Large area (25 mm(2)) silicon drift detectors and detector arrays (5x5) have been designed, simulated, and fabricated for X-ray spectroscopy. On the anode side, the hexagonal drift detector was designed with self-biasing spiral cathode rings (p(+)) of fixed resistance between rings and with a grounded guard anode to separate surface current from the anode current. Two designs have been used for the P-side: symmetric self-biasing spiral cathode rings (p(+)) and a uniform backside p(+) implant. Only 3 to 5 electrodes are needed to bias the detector plus an anode for signal collection. With graded electrical potential, a sub-nanoamper anode current, and a very small anode capacitance, an initial FWHM of 1.3 keV, without optimization of all parameters, has been obtained for 5.9 keV Fe-55 X-ray at RT using a uniform backside detector. |
Identificador | |
Idioma(s) |
英语 |
Fonte |
Zhang WC; Li Z; Siddons DP; Huang T; Zhao LJ; Kakuno EM; Pietraski P; Li CJ .Design, simulation and testing of large area silicon drift detectors and detector array for X-ray spectroscopy ,IEEE TRANSACTIONS ON NUCLEAR SCIENCE,2000,47(4 Part.1.):1381-1385 |
Palavras-Chave | #半导体器件 |
Tipo |
期刊论文 |