At least three known standards are normally required for the full two-port test fixture calibration in vector network analyzers (VNA). In this paper, a calibration procedure using only one standard, based on establishing two hypothetical symmetrical fixtures using triple-through method, is shown. The results using the calibrating method to subtract the influence of fixtures are in accord with the directly measured data of the device-under-test (DUT) without the fixtures very well, which shows that the proposed method is very simple and accurate.

Identificador

http://ir.semi.ac.cn/handle/172111/8698

http://www.irgrid.ac.cn/handle/1471x/63879

Idioma(s)

中文

Fonte

Liu, C; Zhang, SJ; Xie, L; Zhu, NH .A new method based on the construction of hypothetical symmetrical networks for fixture calibration in network analyzers ,ACTA PHYSICA SINICA,JUN 2005,54 (6):2606-2610