激光稳频的共焦法布里-珀罗干涉仪


Autoria(s): 孙旭涛; 刘继桥; 周军; 陈卫标
Data(s)

2008

Resumo

针对多普勒激光雷达激光源短期频率漂移低于1 MHz的要求,设计了一种共焦干涉仪作为频率标准进行稳频。通过对三种不同材料制成的共焦法布里-珀罗(Fabry-Perot)干涉仪中心频率随温度漂移情况进行分析对比,选用零膨胀微晶玻璃材料制作共焦法布里-珀罗干涉仪,腔镜和隔离器通过光胶的方式进行组合,并且置于温控精度优于0.01 K的双层密封温控箱中。经过实验测量,共焦法布里-珀罗干涉仪的自由光谱范围为370 MHz,透射谱半峰全宽(FWHM)为1.7 MHz,精细度为220。采用该共焦干涉仪进行稳频,理论稳频精度可达0.15 MHz,满足激光多普勒雷达单频激光源的稳频要求。

The frequency excursion of laser source of Doppler lidar should be less than 1 MHz. To satisfy this demand, a confocal Fabry-Perot (F-P) interferometer was manufactured as the frequency standard for frequency stabilization. After analyzing and contrasting the center frequency excursion of confocal Fabry-Perot interferometers that are made of three different types of material with the change of temperature, the zerodur material was selected to fabricate the interferometer, and the cavity mirrors were optically contacted onto the end of spacer. The confocal Fabry-Perot interferometer was situated within a double-walled chamber, and the change of temperature in the chamber was less than 0.01 K. The experimental results indicate that the free spectral range is 370 MHz, the full-width at half maximum (FWHM) is 1.7 MHz, and the finesse is 220. Using this interferometer as frequency standard of frequency stabilization, the theoretical frequency excursion is 0.15 MHz, which satisfies the need of frequency stabilization of single mode laser of Doppler lidar.

Identificador

http://ir.siom.ac.cn/handle/181231/2421

http://www.irgrid.ac.cn/handle/1471x/11472

Idioma(s)

中文

Fonte

孙旭涛;刘继桥;周军;陈卫标;.激光稳频的共焦法布里-珀罗干涉仪,中国激光,2008,35(7):1005-1008

Palavras-Chave #激光技术 #激光技术 #共焦法布里-珀罗干涉仪 #自由光谱范围 #激光稳频 #多普勒激光雷达 #Confocal Fabry-Perot interferometer #Doppler lidar #Free spectral range #Frequency stabilization #Laser technique
Tipo

期刊论文