MBARI procedures for automated nutrient analyses using a modified Alpkem Series 300 Rapid Flow Analyzer


Autoria(s): Sakamoto, Carole M.; Friederich, Gernot E.; Codispoti, Louis A.
Data(s)

1990

Resumo

This report documents the methods used at the Monterey Bay Aquarium Research Institute (MBARI) for analyzing seawater nutrient samples with an Alpkem Series 300 Rapid Flow Analyzer (RFA) system. The methods have been optimized for the particular requirements of this laboratory. The RFA system has been used to analyze approximately 20,000 samples during the past two years. The methods have been optimized to run nutrient analyses in a routine manner with a detection limit of better than -±1% and a within run precision of -±1% of the full scale concentration range. The normal concentration ranges are 0-200 ~M silicate, 0-5 ~M phosphate, 0-50 ~M nitrate, 0-3 ~M nitrite, and 0-10 ~M ammonium. The memorandum is designed to be used in a loose-leaf binder format. Each page is dated and as revisions are made, they should be inserted into the binder. The revisions should be added into the binder. Retain the old versions in order to maintain a historical record of the procedures. (88 pages)

Formato

application/pdf

Identificador

http://aquaticcommons.org/1971/1/Technical_Report_90-2.pdf

Sakamoto, Carole M. and Friederich, Gernot E. and Codispoti, Louis A. (1990) MBARI procedures for automated nutrient analyses using a modified Alpkem Series 300 Rapid Flow Analyzer. Moss Landing, CA, Monterey Bay Aquarium Research Institute, (MBARI Technical Report, 90-2)

Idioma(s)

en

Publicador

Monterey Bay Aquarium Research Institute

Relação

http://aquaticcommons.org/1971/

Palavras-Chave #Chemistry
Tipo

Monograph or Serial Issue

NonPeerReviewed