High resolution light-sheet based high-throughput imaging cytometry system enables visualization of intra-cellular organelles
Data(s) |
2014
|
---|---|
Resumo |
Visualization of intracellular organelles is achieved using a newly developed high throughput imaging cytometry system. This system interrogates the microfluidic channel using a sheet of light rather than the existing point-based scanning techniques. The advantages of the developed system are many, including, single-shot scanning of specimens flowing through the microfluidic channel at flow rate ranging from micro-to nano- lit./min. Moreover, this opens-up in-vivo imaging of sub-cellular structures and simultaneous cell counting in an imaging cytometry system. We recorded a maximum count of 2400 cells/min at a flow-rate of 700 nl/min, and simultaneous visualization of fluorescently-labeled mitochondrial network in HeLa cells during flow. The developed imaging cytometry system may find immediate application in biotechnology, fluorescence microscopy and nano-medicine. (C) 2014 Author(s). All article content, except where otherwise noted, is licensed under a Creative Commons Attribution 3.0 Unported License. |
Formato |
application/pdf |
Identificador |
http://eprints.iisc.ernet.in/50271/1/aip_adv_4-9_2014.pdf Regmi, Raju and Mohan, Kavya and Mondal, Partha Pratim (2014) High resolution light-sheet based high-throughput imaging cytometry system enables visualization of intra-cellular organelles. In: AIP ADVANCES, 4 (9). |
Publicador |
AMER INST PHYSICS |
Relação |
http://dx.doi.org/ 10.1063/1.4896260 http://eprints.iisc.ernet.in/50271/ |
Palavras-Chave | #Instrumentation and Applied Physics (Formally ISU) |
Tipo |
Journal Article PeerReviewed |