Blood pressure evaluation using sphygmomanometry assisted by arterial pulse waveform detection by fiber Bragg grating pulse device
Data(s) |
2013
|
---|---|
Resumo |
We report a blood pressure evaluation methodology by recording the radial arterial pulse waveform in real time using a fiber Bragg grating pulse device (FBGPD). Here, the pressure responses of the arterial pulse in the form of beat-to-beat pulse amplitude and arterial diametrical variations are monitored. Particularly, the unique signatures of pulse pressure variations have been recorded in the arterial pulse waveform, which indicate the systolic and diastolic blood pressure while the patient is subjected to the sphygmomanometric blood pressure examination. The proposed method of blood pressure evaluation using FBGPD has been validated with the auscultatory method of detecting the acoustic pulses (Korotkoff sounds) by an electronic stethoscope. (C) 2013 Society of Photo-Optical Instrumentation Engineers (SPIE) |
Formato |
application/pdf |
Identificador |
http://eprints.iisc.ernet.in/47317/1/jou_bio_opt-18_6_2013.pdf Sharath, Umesh and Sukreet, Raju and Apoorva, Girish and Asokan, Sundarrajan (2013) Blood pressure evaluation using sphygmomanometry assisted by arterial pulse waveform detection by fiber Bragg grating pulse device. In: JOURNAL OF BIOMEDICAL OPTICS, 18 (6). |
Publicador |
SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS |
Relação |
http://dx.doi.org/10.1117/1.JBO.18.6.067010 http://eprints.iisc.ernet.in/47317/ |
Palavras-Chave | #Instrumentation and Applied Physics (Formally ISU) |
Tipo |
Journal Article PeerReviewed |