A Compliant End-Effector to Passively Limit the Force in Tele-Operated Tissue-Cutting
Data(s) |
2012
|
---|---|
Resumo |
This paper presents a compliant end-effector that cuts soft tissues and senses the cutting forces. The end-effector is designed to have an upper threshold on cutting forces to facilitate safe handling of tissue during automated cutting. This is demonstrated with nonlinear finite element analysis and experimental results obtained by cutting inhomogeneous phantom tissue. The cutting forces are estimated using a vision-based technique that uses amplified elastic deformation of the compliant end-effector. We also demonstrate an immersive tele-operated tissue-cutting system together with a haptic device that gives real-time force feedback to the user. DOI: 10.1115/1.4007638] |
Formato |
application/pdf |
Identificador |
http://eprints.iisc.ernet.in/45581/1/jol_med_dev_tra_asme_6-4_2012.pdf Bhargav, Santosh DB and Chakravarthy, Shanthanu and Ananthasuresh, GK (2012) A Compliant End-Effector to Passively Limit the Force in Tele-Operated Tissue-Cutting. In: JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 6 (4). |
Publicador |
ASME-AMER SOC MECHANICAL ENG |
Relação |
http://dx.doi.org/10.1115/1.4007638 http://eprints.iisc.ernet.in/45581/ |
Palavras-Chave | #Mechanical Engineering |
Tipo |
Journal Article PeerReviewed |