Development and studies on microstrip detector


Autoria(s): Radhakrishna, V; Rajanna, K; Desai, SS; Shaikh, AM
Data(s)

25/05/2003

Resumo

We report development of gas microstrip detectors using thin film and lithography techniques. The detectors were tested for their performance for X-rays (5.9 keV) and a maximum gas gain of similar to 13,000 and best resolution of similar to 12% was obtained. Factors affecting gain and resolution were investigated. The detectors were tested for their one-dimensional position sensitivity. Meandering resistive strips were used for charge division method. A position resolution of 0.48 mm was obtained.

Formato

application/pdf

Identificador

http://eprints.iisc.ernet.in/39778/1/Development_and.pdf

Radhakrishna, V and Rajanna, K and Desai, SS and Shaikh, AM (2003) Development and studies on microstrip detector. In: Current Science (Bangalore), 84 (10). pp. 1327-1336.

Publicador

Indian Academy of Sciences

Relação

http://www.ias.ac.in/currsci/may252003/contents.htm

http://eprints.iisc.ernet.in/39778/

Palavras-Chave #Instrumentation and Applied Physics (Formally ISU)
Tipo

Journal Article

PeerReviewed