Development and studies on microstrip detector
Data(s) |
25/05/2003
|
---|---|
Resumo |
We report development of gas microstrip detectors using thin film and lithography techniques. The detectors were tested for their performance for X-rays (5.9 keV) and a maximum gas gain of similar to 13,000 and best resolution of similar to 12% was obtained. Factors affecting gain and resolution were investigated. The detectors were tested for their one-dimensional position sensitivity. Meandering resistive strips were used for charge division method. A position resolution of 0.48 mm was obtained. |
Formato |
application/pdf |
Identificador |
http://eprints.iisc.ernet.in/39778/1/Development_and.pdf Radhakrishna, V and Rajanna, K and Desai, SS and Shaikh, AM (2003) Development and studies on microstrip detector. In: Current Science (Bangalore), 84 (10). pp. 1327-1336. |
Publicador |
Indian Academy of Sciences |
Relação |
http://www.ias.ac.in/currsci/may252003/contents.htm http://eprints.iisc.ernet.in/39778/ |
Palavras-Chave | #Instrumentation and Applied Physics (Formally ISU) |
Tipo |