Flow evaluation and hemolysis analysis of BVAD centrifugal blood pump by computational fluids dynamics


Autoria(s): Bumrungpetch, Jeerasit; Tan, Andy Chit; Liu, Shu-Hong; Luo, Xian-Wu; Wu, Qing-Yu; Yuan, Jian-Ping; Zhang, Ming-Kui
Data(s)

2014

Resumo

Computational fluid dynamics (CFD) and particle image velocimetry (PIV) are commonly used techniques to evaluate the flow characteristics in the development stage of blood pumps. CFD technique allows rapid change to pump parameters to optimize the pump performance without having to construct a costly prototype model. These techniques are used in the construction of a bi-ventricular assist device (BVAD) which combines the functions of LVAD and RVAD in a compact unit. The BVAD construction consists of two separate chambers with similar impellers, volutes, inlet and output sections. To achieve the required flow characteristics of an average flow rate of 5 l/min and different pressure heads (left – 100mmHg and right – 20mmHg), the impellers were set at different rotating speeds. From the CFD results, a six-blade impeller design was adopted for the development of the BVAD. It was also observed that the fluid can flow smoothly through the pump with minimum shear stress and area of stagnation which are related to haemolysis and thrombosis. Based on the compatible Reynolds number the flow through the model was calculated for the left and the right pumps. As it was not possible to have both the left and right chambers in the experimental model, the left and right pumps were tested separately.

Identificador

http://eprints.qut.edu.au/88648/

Publicador

Yuce Gi'gye Gong'eob Haghoe / Korean Fluid Machinery Association

Relação

DOI:10.5293/IJFMS.2014.7.1.034

Bumrungpetch, Jeerasit, Tan, Andy Chit, Liu, Shu-Hong, Luo, Xian-Wu, Wu, Qing-Yu, Yuan, Jian-Ping, & Zhang, Ming-Kui (2014) Flow evaluation and hemolysis analysis of BVAD centrifugal blood pump by computational fluids dynamics. International Journal of Fluid Machinery and Systems, 7(1), pp. 34-41.

Fonte

School of Chemistry, Physics & Mechanical Engineering; Science & Engineering Faculty

Palavras-Chave #Blood Pump, Ventricular Assist device, LVAD, BVAD, Computational fluid dynamics, artificial heart
Tipo

Journal Article