Field size consistency of nominally matched linacs


Autoria(s): Kairn, Tanya; Asena, Andre; Charles, Paul; Hill, Brendan; Langton, Christian M.; Middlebrook, Nigel; Moylan, Rebecca; Trapp, Jamie
Data(s)

15/05/2015

Resumo

Given that there is increasing recognition of the effect that submillimetre changes in collimator position can have on radiotherapy beam dosimetry, this study aimed to evaluate the potential variability in small field collimation that may exist between otherwise matched linacs. Field sizes and field output factors were measured using radiochromic film and an electron diode, for jaw- and MLC-collimated fields produced by eight dosimetrically matched Varian iX linacs (Varian Medical Systems, Palo Alto, USA). This study used nominal sizes from 0.6×0.6 to 10×10 cm215 , for jaw-collimated fields,and from 1×1 to 10×10 cm216 , for MLC-collimated fields, delivered from a zero (head up, beam directed vertically downward) gantry angle. Differences between the field sizes measured for the eight linacs exceeded the uncertainty of the film measurements and the repositioning uncertainty of the jaws and MLCs on one linac. The dimensions of fields defined by MLC leaves were more consistent between linacs, while also differing more from their nominal values than fields defined by orthogonal jaws. The field output factors measured for the different linacs generally increased with increasing measured field size for the nominal 0.6×0.6 and 1×1 cm2 fields, and became consistent between linacs for nominal field sizes of 2×2 cm2 25 and larger. The inclusion in radiotherapy treatment planning system beam data of small field output factors acquired in fields collimated by jaws (rather than the more-reproducible MLCs), associated with either the nominal or the measured field sizes, should be viewed with caution. The size and reproducibility of the fields (especially the small fields) used to acquire treatment planning data should be investigated thoroughly as part of the linac or planning system commissioning process. Further investigation of these issues, using different linac models, collimation systems and beam orientations, is recommended.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/84523/

Publicador

Springer

Relação

http://eprints.qut.edu.au/84523/1/__staffhome.qut.edu.au_staffgroupt%24_trapp_Documents_Jamie_Research_eprints_kairn_fieldsizeR2.pdf

DOI:10.1007/s13246-015-0349-2

Kairn, Tanya, Asena, Andre, Charles, Paul, Hill, Brendan, Langton, Christian M., Middlebrook, Nigel, Moylan, Rebecca, & Trapp, Jamie (2015) Field size consistency of nominally matched linacs. Australasian Physical and Engineering Sciences in Medicine. (In Press)

http://purl.org/au-research/grants/ARC/LP110100401

Direitos

Copyright 2015 Springer

Fonte

School of Chemistry, Physics & Mechanical Engineering; Science & Engineering Faculty

Palavras-Chave #029903 Medical Physics #linear accelerator #multi-leaf collimator #radiation therapy