Is there a bone-nail specific entry point? automated fit quantification of tibial nail designs during the insertion for six different nail entry points


Autoria(s): Amarathunga, Jayani P.; Schuetz, Michael; Yarlagadda, K. D. V.; Schmutz, Beat
Data(s)

01/04/2015

Resumo

Intramedullary nailing is the standard fixation method for displaced diaphyseal fractures of tibia. Selection of the correct nail insertion point is important for axial alignment of bone fragments and to avoid iatrogenic fractures. However, the standard entry point (SEP) may not always optimise the bone-nail fit due to geometric variations of bones. This study aimed to investigate the optimal entry for a given bone-nail pair using the fit quantification software tool previously developed by the authors. The misfit was quantified for 20 bones with two nail designs (ETN and ETN-Proximal Bend) related to the SEP and 5 entry points which were 5 mm and 10 mm away from the SEP. The SEP was the optimal entry point for 50% of the bones used. For the remaining bones, the optimal entry point was located 5 mm away from the SEP, which improved the overall fit by 40% on average. However, entry points 10 mm away from the SEP doubled the misfit. The optimised bone-nail fit can be achieved through the SEP and within the range of a 5 mm radius, except posteriorly. The study results suggest that the optimal entry point should be selected by considering the fit during insertion and not only at the final position.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/80832/

Publicador

Elsevier Ltd

Relação

http://eprints.qut.edu.au/80832/1/Is%20there%20a%20bone-nail%20specific%20entry%20point_Automated%20fit%20quantification%20of%20tibial%20nail%20designs%20during%20the%20insertion%20for%20six%20different%20nail%20entry%20points_eprints.pdf

DOI:10.1016/j.medengphy.2015.01.012

Amarathunga, Jayani P., Schuetz, Michael, Yarlagadda, K. D. V., & Schmutz, Beat (2015) Is there a bone-nail specific entry point? automated fit quantification of tibial nail designs during the insertion for six different nail entry points. Medical Engineering & Physics, 37(4), pp. 367-374.

Direitos

Copyright 2015 Elsevier Ltd

NOTICE: this is the author’s version of a work that was accepted for publication in Medical Engineering & Physics. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. A definitive version was subsequently published in Medical Engineering & Physics, Volume 37, Issue 4, (April 2015), DOI: 10.1016/j.medengphy.2015.01.012

Fonte

School of Chemistry, Physics & Mechanical Engineering; Institute of Health and Biomedical Innovation; Science & Engineering Faculty

Palavras-Chave #Intramedullary nail #nail fit #nail insertion point #tibia #bone-nail specific insertion point