Measurements of half-field beam penumbra were taken using EBT2 film for a variety of blocking techniques. It was shown that minimizing the SSD reduces the penumbra as the effects of beam divergence are diminished. The addition of a lead block directly on the surface provides optimal results with a 10-90% penumbra of 0.53 ± 0.02 cm. To resolve the uncertainties encountered in film measurements, future Monte Carlo measurements of halffield penumbras are to be conducted. |
Formato |
application/pdf |
Identificador | |
Publicador |
Institute of Physics Publishing Ltd. |
Relação |
http://eprints.qut.edu.au/80030/1/__staffhome.qut.edu.au_staffgrouph%24_hollambc_Desktop_1742-6596_573_1_012073.pdf DOI:10.1088/1742-6596/573/1/012073 Smith, Shaun, Kairn, Tanya, Crowe, Scott, Asena, Andre, & Trapp, Jamie (2015) Characterisation of the half-field beam penumbra for a variety of blocking set-ups. Journal of Physics: Conference Series, 573(1), 012073. http://purl.org/au-research/grants/ARC/LP110100401 WESLEY RESEARCH INST/LP110100401 |