Internal calibration of gel dosimeters : a feasibility study
Data(s) |
2009
|
---|---|
Resumo |
In this work we test the feasibility of a new calibration method for gel dosimetry. We examine, through Monte Carlo modelling, whether the inclusion of an organic plastic scintillator system at key points within the gel phantom would perturb the dose map. Such a system would remove the requirement for a separate calibration gel, removing many sources of uncertainty. |
Formato |
application/pdf |
Identificador | |
Publicador |
Institute of Physics Publishing Ltd. |
Relação |
http://eprints.qut.edu.au/75333/1/75333.pdf http://iopscience.iop.org/1742-6596/164/1/012014 DOI:10.1088/1742-6596/164/1/012014 Trapp, Jamie, Kairn, Tanya, Crowe, Scott, & Fielding, Andrew (2009) Internal calibration of gel dosimeters : a feasibility study. Journal of Physics: Conference Series, 164(1), 012014. |
Direitos |
Copyright 2009 IOP Publishing Ltd Published under licence in Journal of Physics: Conference Series by IOP Publishing Ltd. Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. |
Fonte |
School of Chemistry, Physics & Mechanical Engineering; Science & Engineering Faculty |
Palavras-Chave | #029903 Medical Physics |
Tipo |
Journal Article |