Resumo

Two sources of uncertainty in the X ray computed tomography imaging of polymer gel dosimeters are investigated in the paper.The first cause is a change in postirradiation density, which is proportional to the computed tomography signal and is associated with a volume change. The second cause of uncertainty is reconstruction noise.A simple technique that increases the residual signal to noise ratio by almost two orders of magnitude is examined.

Identificador

http://eprints.qut.edu.au/70190/

Publicador

IAEA

Relação

http://www-pub.iaea.org/MTCD/publications/PDF/Pub1153/CD/P1153_2.pdf

Trapp, Jamie, Michael, Gregory, De Deene, Yves, & Baldock, Clive (2003) Factors affecting the extraction of absorbed dose information in a 3D polymer gel dosimeter by X ray computed tomography (IAEA-CN-96/122P). In Standards and Codes of Practice in Medical Radiation Dosimetry, IAEA, 25–28 November 2002, pp. 389-396.