Reliability of spatiotemporal and kinetic gait parameters determined by a new instrumented treadmill system


Autoria(s): Reed, Lloyd F.; Urry, Stephen R.; Wearing, Scott C.
Data(s)

14/08/2013

Resumo

Background Despite the emerging use of treadmills integrated with pressure platforms as outcome tools in both clinical and research settings, published evidence regarding the measurement properties of these new systems is limited. This study evaluated the within– and between–day repeatability of spatial, temporal and vertical ground reaction forces measured by a treadmill system instrumented with a capacitance–based pressure platform. Methods Thirty three healthy adults (mean age, 21.5 ± 2.8 years; height, 168.4 ± 9.9 cm; and mass, 67.8 ± 18.6 kg), walked barefoot on a treadmill system (FDM–THM–S, Zebris Medical GmbH) on three separate occasions. For each testing session, participants set their preferred pace but were blinded to treadmill speed. Spatial (foot rotation, step width, stride and step length), temporal (stride and step times, duration of stance, swing and single and double support) and peak vertical ground reaction force variables were collected over a 30–second capture period, equating to an average of 52 ± 5 steps of steady–state walking. Testing was repeated one week following the initial trial and again, for a third time, 20 minutes later. Repeated measures ANOVAs within a generalized linear modelling framework were used to assess between–session differences in gait parameters. Agreement between gait parameters measured within the same day (session 2 and 3) and between days (session 1 and 2; 1 and 3) were evaluated using the 95% repeatability coefficient. Results There were statistically significant differences in the majority (14/16) of temporal, spatial and kinetic gait parameters over the three test sessions (P < .01). The minimum change that could be detected with 95% confidence ranged between 3% and 17% for temporal parameters, 14% and 33% for spatial parameters, and 4% and 20% for kinetic parameters between days. Within–day repeatability was similar to that observed between days. Temporal and kinetic gait parameters were typically more consistent than spatial parameters. The 95% repeatability coefficient for vertical force peaks ranged between ± 53 and ± 63 N. Conclusions The limits of agreement in spatial parameters and ground reaction forces for the treadmill system encompass previously reported changes with neuromuscular pathology and footwear interventions. These findings provide clinicians and researchers with an indication of the repeatability and sensitivity of the Zebris treadmill system to detect changes in common spatiotemporal gait parameters and vertical ground reaction forces.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/65420/

Publicador

BioMed Central Ltd.

Relação

http://eprints.qut.edu.au/65420/1/QUT_EPRINTS_BMC_RELIABILITY_Formatted_scw.pdf

http://www.biomedcentral.com/1471-2474/14/249

DOI:10.1186/1471-2474-14-249

Reed, Lloyd F., Urry, Stephen R., & Wearing, Scott C. (2013) Reliability of spatiotemporal and kinetic gait parameters determined by a new instrumented treadmill system. BMC Musculoskeletal Disorders, 14(249).

Direitos

Copyright 2013 Reed et al.; licensee BioMed Central Ltd.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Fonte

School of Clinical Sciences; Faculty of Health; Institute of Health and Biomedical Innovation

Palavras-Chave #090302 Biomechanical Engineering #110318 Podiatry #110601 Biomechanics #110604 Sports Medicine #Gait analysis #Walking #Measurement #Repeatability #Precision #Reproducibility
Tipo

Journal Article