Measurement of a PAGAT gel dosimeter by ultrasound computed tomography


Autoria(s): Khoei, Shadi; Trapp, Jamie; Langton, Christian M.
Data(s)

2013

Resumo

In this work we used a 3D quantitative CT ultrasound imaging system to characterise polymer gel dosimeters. The system comprised of two identical 5 MHz 128 element phased-array ultrasound transducers co-axially aligned and submerged in water as a coupling agent. Rotational and translational movement of the gel dosimeter sample between the transducers were performed using a robotic arm. Ultrasound signals were generated and received using an Olympus Omniscan unit. Dose sensitivity of attenuation and time of flight ultrasonic parameters were assessed using this system.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/61029/

Publicador

Insitute of Physics Publishing

Relação

http://eprints.qut.edu.au/61029/1/Khoei_JPCS_2013_UCT.pdf

DOI:10.1088/1742-6596/444/1/012083

Khoei, Shadi, Trapp, Jamie, & Langton, Christian M. (2013) Measurement of a PAGAT gel dosimeter by ultrasound computed tomography. Journal of Physics: Conference Series, 444, 012083.

Direitos

Copyright 2013 (please consult the author).

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by IOP Publishing Ltd