Monte Carlo modeling of a 4 mm conical collimator for a Novalis Tx Linear Accelerator


Autoria(s): Morales, Johnny; Crowe, Scott; Trapp, Jamie
Data(s)

2012

Resumo

The work presented in this poster outlines the steps taken to model a 4 mm conical collimator (BrainLab, Germany) on a Novalis Tx linear accelerator (Varian, Palo Alto, USA) capable of producing a 6MV photon beam for treatment of Stereotactic Radiosurgery (SRS) patients. The verification of this model was performed by measurements in liquid water and in virtual water. The measurements involved scanning depth dose and profiles in a water tank plus measurement of output factors in virtual water using Gafchromic® EBT3 film.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/50682/

Relação

http://eprints.qut.edu.au/50682/1/Sevilla_2012_poster_Morales.pdf

http://www.mctp2012.com/

Morales, Johnny, Crowe, Scott, & Trapp, Jamie (2012) Monte Carlo modeling of a 4 mm conical collimator for a Novalis Tx Linear Accelerator. In EWGMCTP - Third European Workshop on Monte Carlo Treatment Planning, 15-18 May 2012, Seville, Spain.

Direitos

Copyright 2012 (please consult the authors).

Fonte

School of Chemistry, Physics & Mechanical Engineering; Faculty of Science and Technology

Palavras-Chave #029900 OTHER PHYSICAL SCIENCES #029903 Medical Physics #Monte Carlo #Stereotactic #radiosurgery #Novalis #radiotherapy