Flexural and torsional rigidity of colonoscope at room and body temperatures


Autoria(s): Hellier, David; Albermani, Faris; Evans, Brendon; de Visser, Hans; Adam, Clayton J.; Passenger, Joshua D.; ,
Data(s)

01/04/2011

Resumo

A series of experiments have been conducted to determine the flexural, EI, and torsional, GJ, rigidity of an Olympus colonoscope CF‐140S and torsional rigidity of a Pentax colonoscope EC‐3870 and the dependency of these properties on temperature and on the presence of loops. Along the length of the colonoscope, the Olympus colonoscope flexural rigidity varied between 260 and 400 Ncm2 and torsional rigidity varied between 68 and 88 Ncm2/deg, with an average of 76 Ncm2/deg for tests involving 0.86 Nm of anti‐clockwise torque. Results show a significant decrease of 10% in torsional rigidity between clockwise and anti‐clockwise torque. For the Pentax colonoscope flexural rigidity was not tested; its torsional rigidity varied between 34 and 76 Ncm2/deg, with an average of 46 Ncm2/deg for tests involving 0.43 Nm of anti‐clockwise torque. An increase in temperature of the Olympus colonoscope from 24°C to 37°C reduces EI by an average of 17% and GJ by an average of 7%. A right‐handed loop caused a significant increase in flexural rigidity, but other looping configurations had no significant influence.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/42392/

Publicador

Sage Publications Ltd.

Relação

http://eprints.qut.edu.au/42392/1/389.full.pdf

DOI:10.1177/09544119JEIM883

Hellier, David, Albermani, Faris, Evans, Brendon, de Visser, Hans, Adam, Clayton J., Passenger, Joshua D., & , (2011) Flexural and torsional rigidity of colonoscope at room and body temperatures. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 225(4), pp. 389-399.

Direitos

Copyright 2011 Sage Publications Ltd.

Fonte

Faculty of Built Environment and Engineering; Institute of Health and Biomedical Innovation; School of Engineering Systems

Palavras-Chave #090302 Biomechanical Engineering #torsional rigidity #flexural rigidity #colonoscope mechanical properties #effect of temperature on mechanical properties #colonoscope
Tipo

Journal Article