Investigation of stereotactic radiotherapy dose using dosimetry film and Monte Carlo simulations


Autoria(s): Kairn, Tanya; Crowe, Scott; Kenny, John; Trapp, Jamie
Data(s)

17/06/2011

Resumo

This study uses dosimetry film measurements and Monte Carlo simulations to investigate the accuracy of type-a (pencil-beam) dose calculations for predicting the radiation doses delivered during stereotactic radiotherapy treatments of the brain. It is shown that when evaluating doses in a water phantom, the type-a algorithm provides dose predictions which are accurate to within clinically relevant criteria, gamma(3%,3mm), but these predictions are nonetheless subtly different from the results of evaluating doses from the same fields using radiochromic film and Monte Carlo simulations. An analysis of a clinical meningioma treatment suggests that when predicting stereotactic radiotherapy doses to the brain, the inaccuracies of the type-a algorithm can be exacerbated by inadequate evaluation of the effects of nearby bone or air, resulting in dose differences of up to 10% for individual fields. The results of this study indicate the possible advantage of using Monte Carlo calculations, as well as measurements with high-spatial resolution media, to verify type-a predictions of dose delivered in cranial treatments.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/42244/

Publicador

Elsevier BV

Relação

http://eprints.qut.edu.au/42244/1/kairn_RadMeas_Stereo_2011.pdf

DOI:10.1016/j.radmeas.2011.06.024

Kairn, Tanya, Crowe, Scott, Kenny, John, & Trapp, Jamie (2011) Investigation of stereotactic radiotherapy dose using dosimetry film and Monte Carlo simulations. Radiation Measurements, 46(12), pp. 1985-1988.

http://purl.org/au-research/grants/arc/LP110100401

Direitos

Copyright 2011 Elsevier

NOTICE: this is the author’s version of a work that was accepted for publication in Radiation Measurements. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. A definitive version was subsequently published in Radiation Measurements, [VOL 46, ISSUE 12, (2011)] 10.1016/j.radmeas.2011.06.024

Fonte

Faculty of Science and Technology; Physics

Palavras-Chave #029900 OTHER PHYSICAL SCIENCES #029903 Medical Physics #computer simulation #dosimetry #pencil beam #radiochromic film #stereotactic radiosurgery #radiotherapy #Monte Carlo #BEAM #radiation therapy #treatment planning
Tipo

Journal Article