A comparative study of the effect of calibration conditions on the water equivalence of a range of gel dosimeters


Autoria(s): Taylor, Michael; Franich, Rick; Trapp, Jamie; Johnston, Peter
Data(s)

01/04/2009

Resumo

Gel dosimeters are of increasing interest in the field of radiation oncology as the only truly three-dimensional integrating radiation dosimeter. There are a range of ferrous-sulphate and polymer gel dosimeters. To be of use, they must be water-equivalent. On their own, this relates to their radiological properties as determined by their composition. In the context of calibration of gel dosimeters, there is the added complexity of the calibration geometry; the presence of containment vessels may influence the dose absorbed. Five such methods of calibration are modelled here using the Monte Carlo method. It is found that the Fricke gel best matches water for most of the calibration methods, and that the best calibration method involves the use of a large tub into which multiple fields of different dose are directed. The least accurate calibration method involves the use of a long test tube along which a depth dose curve yields multiple calibration points.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/37733/

Publicador

Institute of Electrical and Electronics Engineers

Relação

http://eprints.qut.edu.au/37733/1/c37733.pdf

DOI:10.1109/TNS.2009.2015315

Taylor, Michael, Franich, Rick, Trapp, Jamie, & Johnston, Peter (2009) A comparative study of the effect of calibration conditions on the water equivalence of a range of gel dosimeters. IEEE Transactions on Nuclear Science, 56(2), pp. 429-436.

Direitos

Copyright 2009 IEEE

Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE.

Fonte

Faculty of Science and Technology; Physics

Palavras-Chave #029903 Medical Physics #Radiotherapy #Gel dosimetry #Monte Carlo
Tipo

Journal Article