Microwave decontamination of eyelid warming devices for the treatment of meibomian gland dysfunction


Autoria(s): Bilkhu, Paramdeep; Wolffsohn, James S.; Hilton, Anthony; Matthews, Gabrielle; Legood, Simon; Riley, Liam
Resumo

PURPOSE: The role of bacteria in meibomian gland dysfunction is unclear, yet contamination of compresses used as treatment may exacerbate this condition. This study therefore determined the effect of heating on bacteria on two forms of compress. METHODS: Cotton flannels and MGDRx EyeBags (eyebags) were inoculated by adding experimental inoculum (Staphylococcus aureus, Streptococcus pyogenes, Pseudomonas aeruginosa; one species for each set of 3 eyebags and flannels). One of each were then randomised in to 3 groups: no heating (control); therapeutic (47.4±0.7°C); or sanitisation (68±1.1°C). After treatment, bacteria cell numbers were calculated. The experiment was repeated in triplicate. RESULTS: There was a statistically significant difference between each treatment with the eyebag for S. aureus (control=7.15±0.11logC/ml, therapeutic heating=5.24±0.59logC/ml, sanitisation heating=3.48±1.43logC/ml; P<0.001) and S. pyogenes (7.36±0.13, 5.73±0.26, 4.75±0.54; P<0.001). P. aeruginosa also showed a significant reduction (P<0.001) from control (6.39±0.34) to therapeutic (0.33±0.26) and sanitisation (0.33±0.21), but the latter were similar (P=1.000). For the flannels, there was significant difference between each treatment for S. aureus (6.89±0.46, 3.96±1.76, 0.42±0.90; P<0.001). For S. pyogenes, there was a significant reduction (P<0.001) from control (7.51±0.10) to therapeutic (5.91±0.62) and sanitisation (5.18±0.8), but the latter were similar (P=0.07). For P. aeruginosa, there was a significant difference (P<0.001) from control (7.15±0.36) to sanitisation (5.83±0.44); but not to therapeutic (6.84±0.31) temperatures (P=0.07). CONCLUSIONS: Therapeutic heating produces a significant reduction in bacteria on the eyebags, but only sanitisation heating appears effective for flannels. However, patients should be advised to heat the eyebag to sanitisation temperatures on initial use.

Formato

application/pdf

Identificador

http://publications.aston.ac.uk/27470/1/Treatment_of_meibomian_gland_dysfunction.pdf

Bilkhu, Paramdeep; Wolffsohn, James S.; Hilton, Anthony; Matthews, Gabrielle; Legood, Simon and Riley, Liam Microwave decontamination of eyelid warming devices for the treatment of meibomian gland dysfunction. Contact Lens and Anterior Eye, 39 (4), 293–297.

Relação

http://publications.aston.ac.uk/27470/

Direitos

cc_by_nc_nd

Tipo

Article

PeerReviewed