Modelling, simulation and characterization of epithelial cell culture biochip


Autoria(s): Candito, Antonio
Contribuinte(s)

Tartagni, Marco

Data(s)

19/03/2015

Resumo

A microfluidic Organ-on-Chip has been developed for monitoring the epithelial cells monolayer. Equivalent circuit Model was used to determine the electrical properties from the impedance spectra of the epithelial cells monolayer. Black platinum on platinum electrodes was electrochemically deposited onto the surface of electrodes to reduce the influence of the electrical double layer on the impedance measurements. Measurements of impedance with an Impedance Analyzer were done to validate the equivalent circuit model and the decrease of the double layer effect. A Lock-in Amplifier was designed to measure the impedance.

Formato

application/pdf

Identificador

http://amslaurea.unibo.it/8483/1/Candito_Antonio_Tesi.pdf

Candito, Antonio (2015) Modelling, simulation and characterization of epithelial cell culture biochip. [Laurea magistrale], Università di Bologna, Corso di Studio in Ingegneria biomedica [LM-DM270] - Cesena <http://amslaurea.unibo.it/view/cds/CDS8198/>

Relação

http://amslaurea.unibo.it/8483/

Direitos

info:eu-repo/semantics/openAccess

Palavras-Chave #organ, on, chip, tight, junctions, double, layer, effect, impedance, lock, in, amplifer #scuola :: 843884 :: Ingegneria e Architettura #cds :: 8198 :: Ingegneria biomedica [LM-DM270] - Cesena #sessione :: terza